Model Number BI-700-00027-120
Device Problems Computer Software Problem (1112); Imprecision (1307); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The representative scanned right and left sides of the spine model with 2 separate navigation's system without issue.It was not possible to duplicate the reported issue.No parts were replaced or returned to the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while in a spinal fusion procedure, the imaging system was showing an alleged inaccuracy of approximately 4mm after taking a spin.The site then brought in another imaging system, took a second spin, and confirmed accuracy.There was a reported delay to the procedure of less than 1 hour due to this issue and no impact on the patient outcome.
 
Manufacturer Narrative
The event reported in this 3500a also represents a potential accidental radiation occurrence (aro) per 21 cfr 1002.20(a).Per 21 cfr 1002.20(c), this event is being reported under part 803.(b)(6).Number of persons exposed: 1.Number of persons adversely affected: 0.Actions taken to control, correct, or eliminate: see narrative above and/or previous mdr submissions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7224697
MDR Text Key98539605
Report Number1723170-2018-00432
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physicist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/04/2018
Initial Date FDA Received01/29/2018
Supplement Dates Manufacturer Received01/24/2018
Supplement Dates FDA Received02/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age67 YR
Patient Weight113
-
-