• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE AIR 40/64/UCT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE AIR 40/64/UCT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Device Problems Device Operates Differently Than Expected (2913); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
We have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
The customer reported that there was noise in the speaker when speaking with the patient through the ct box.The philips field service engineer (fse) confirmed that the operator could only hear the patient intermittently due to the noise coming from the ct box speaker.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This event is currently under investigation.
 
Manufacturer Narrative
The customer reported that the operator could only hear the patient intermittently due to the noise coming from the ct box speaker.The customer confirmed with the philips helpdesk that there was no patient impact and no harm as a result of this event.The philips field service engineer was dispatched to the customer site and evaluated the ct system and determined that there was noise in the speaker when the operator was speaking with patient through the ct box.The fse replaced the ct box assembly, gantry microphone assembly and speaker to correct and resolve the malfunction.The system is operational and in clinical use.Further investigation of this event has determined that it is not reportable.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE AIR 40/64/UCT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
Manufacturer Contact
derrick massuri
595 miner rd
cleveland, OH 44143
4404833000
MDR Report Key7225268
MDR Text Key98640989
Report Number1525965-2018-00091
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeIS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/03/2018
Initial Date FDA Received01/29/2018
Supplement Dates Manufacturer Received01/03/2018
Supplement Dates FDA Received08/14/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-