• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Rash (2033)
Event Type  Injury  
Manufacturer Narrative
Analysis: the samples were not returned; therefore, evaluation was not able to be performed.A lot history review and a device history record (dhr) review could not be conducted because product identifiers were unavailable.In reviewing the labeling supplied with this product, it was found that the instructions for use (ifu) sufficiently address these potential factors.The ifu states ¿potential adverse events that may be unique to the lutonix catheter paclitaxel drug coating include: allergic reaction to drug coating.There may be other potential adverse events that are unforeseen at this time.¿ conclusion: the actual samples were not received for evaluation.Although requested, a lot history review and a device history record (dhr) review could not be conducted because product identifiers were unavailable.The investigation was limited to the literature article and the follow up information obtained through field inquires.It is unknown if patient physiology or an adverse reaction(s) with other devices/medications from the procedure could have contributed to the event.The patient was treated with a topical ointment and oral steroids, then released from the hospital.A review of our quality database noted one additional event occur from the same hospital and operating physician.Lutonix continues to monitor the field performance of this device to detect similar events, should they occur.If additional information is obtained from the field, a supplemental report will be submitted with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported through the journal "jaad case reports, volume 4, number 1", two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters allegedly resulted in cutaneous polyarteritis nodosa of the left leg.The patient developed cutaneous polyarteritis nodosa, or rash, in approximately 4 weeks after the index procedure.As a result, the patient was hospitalized and diagnostic testing was performed involving a punch biopsy of a nodule.The health care professional (hcp) examined the patient, who presented with painful erythematous nodules extending from the left ankle to the mid thigh.Necrosis, ulcers, and livedo reticularis were not present.Sensory and motor examinations were unremarkable, with no signs of neuropathy.The patient was treated with topical ointment and oral prednisone, then released from the hospital.Although requested, no additional information has been obtained from the author of the case report.It is unknown if additional adverse patient effects were reported.The lutonix dcb was discarded by the user facility and not available for return.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00014.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Type of Device
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
john risse
9409 science center dr
new hope, MN 55428
7634632917
MDR Report Key7226198
MDR Text Key98494052
Report Number3006513822-2018-00013
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature,other
Reporter Occupation Physician
Type of Report Initial
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9004
Device Catalogue NumberUNK SFA 035
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age59 YR
-
-