• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF283
Device Problem Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the balloon "looked strange" during inflation.The case was completed with cryo.After the procedure, the inner balloon was observed to be kinked.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the balloon catheter (2af283 with lot: 43671) was returned and analyzed.Visual inspection of the catheter showed that the catheter was intact with no apparent issue.Smart chip verification showed that the catheter had been used for eighteen injections.No system notices were received during performance testing.The dissection showed that the guide wire lumen was kinked on 1.4025 inches from the tip of the catheter.The catheter failed the performance test due to the kink on the guide wire lumen.In conclusion, the reported issue was confirmed through testing.The reported catheter failed the returned product inspection due to a guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7230629
MDR Text Key98740178
Report Number3002648230-2018-00055
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/27/2019
Device Model Number2AF283
Device Catalogue Number2AF283
Device Lot Number43671
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2018
Initial Date FDA Received01/31/2018
Supplement Dates Manufacturer Received02/28/2018
Supplement Dates FDA Received03/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-