MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (CE) ILAB ULTRASOUND IMAGING SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (CE) ILAB ULTRASOUND IMAGING SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number H749ILAB100C270
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: device analysis revealed no visible external damage or defects observed.The unit meets specifications of mdu-5plus basic functional test.The device did not exhibit any malfunction.The returned device passed all performance criteria during analysis.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event (b)(4).
 
Event Description
Same case as: 2134265-2018-01002 and 2134265-2018-01003.It was reported that automatic pullback failure occurred.A 100v ilab ultrasound imaging system was used in conjunction with an imaging catheter and a pullback sled to view the target lesion.During the procedure, pullback was performed, however, it froze for 5 to 6 seconds.Reboot was performed but the issue was not resolved.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILAB ULTRASOUND IMAGING SYSTEM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7233885
MDR Text Key98861438
Report Number2134265-2018-00668
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749ILAB100C270
Device Catalogue NumberILAB100C27
Device Lot Number10036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2018
Initial Date FDA Received02/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-