Model Number SC-8216-70 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/28/2017 |
Event Type
Injury
|
Event Description
|
A report was received that the patient had a previous explant wherein the physician cut the lead but was unsure if the remaining lead was still left inside the patients body.
|
|
Manufacturer Narrative
|
Additional information was received that imaging was done and showed that part of the paddle was still remained in the patients body.The patient underwent a revision procedure wherein the cut part of the paddle was removed and new leads were implanted.No device malfunction was suspected.The explanted device was not returned to bsn as it was discarded by the medical facility.
|
|
Event Description
|
A report was received that the patient had a previous explant wherein the physician cut the lead but was unsure if the remaining lead was still left inside the patients body.
|
|