Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Analysis revealed no damage to the capio slim suture capturing device.The needle dart was retrieved from behind the cage.There is no amount of suture attached to the needle indicating that the needle pulled off of the suture.The uphold mesh assembly was not returned.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Review and analysis of all available information fails to indicate a probable cause.Therefore, the assigned complaint investigation conclusion code for this event is undeterminable.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a cystocele repair procedure on (b)(6) 2017.According to the complainant, during the procedure, the needle did not penetrate the tissue when the capio slim was actuated to implant the mesh leg on the right side.The procedure was completed with another uphold lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.This event has been deemed reportable based on the investigation results: the needle dart was retrieved from behind the cage of the suturing device.There is no amount of suture attached to the needle indicating that the needle pulled off of the suture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7236096
MDR Text Key98997572
Report Number3005099803-2018-00100
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/05/2020
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000058083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2018
Initial Date FDA Received02/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
-
-