Product event summary: the data files for the date of the reported event and balloon catheter, 2af283 with lot 30381, were returned and analyzed.The data files showed two unrelated system notices were received at the date of the case.Visual inspection showed that the balloon catheter was intact with no apparent issues.The id of the push button luer was equal to 0.045 inches.Visual inspection under the microscope did not shows any debris at the proximal end of the guide wire lumen shaft and the luer.The proximal end of the shaft was perfectly circular.No blockage was observed.The catheter passed the deflection test as per specification.The dissection testing confirmed a guide wire lumen kink on 1.389 inches from the distal point of the catheter¿s tip.The reported issue (balloon shape and debris in balloon) was confirmed through testing but not confirmed through data analysis.In conclusion, the balloon catheter, 2af283 with lot 30381, failed the returned product inspection due to a guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the balloon shape was not round and difficult to position.Additionally, and unknown system notice was received indicating the balloon catheter needed to be replaced.It was noted that a filament was visible at the center of the balloon.The balloon catheter was replaced.The case was completed with cryo.No patient complications have been reported as a result of this event.Incoming information indicated that it was the injection tube of the balloon catheter that was observed in the center of the balloon.The "filament" that was observed was the balloon catheter kinked due to difficult manipulation.The balloon catheter subsequently tested out of specification per the manufacturer's investigation.
|