• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE Back to Search Results
Model Number 055
Device Problems Thermal Decomposition of Device (1071); Fire (1245)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer stated "the cord caught fire" the product was returned for investigation.The product was approx.22 years and 3 months old when the incident occurred.This pad was manufactured in 1994.Upon further investigation into the customers complaint, the inspector found that the cord had a small burn right by the strain relief.This is likely due to excessive bending of the cord over an extended period to time.Based on the bce review of feedbacks, bce did not observe a similar issue within the lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE
Type of Device
THERMOPHORE
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
702 s reed rd
fremont IN 46737
Manufacturer Contact
uday bahal
MDR Report Key7241178
MDR Text Key99188008
Report Number1832415-2017-07333
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number055
Device Lot Number943000
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/05/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-