Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.As such, the investigation will be closed.If the complaint device is received in the future we will reopen the complaint and perform the investigation as appropriate.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
It was reported that a patient underwent an atrial fibrillation procedure with a smartablate irrigation tube set and foreign material was discovered.During the procedure, when the tubing set was taken from the package, foreign matter was contained in the tube.Priming was conducted but the issue continued.The issue was resolved by changing the tubing set to another one.This tubing set was not used on the patient.The procedure was completed with no patient consequence.This event is mdr reportable because if foreign material is found inside the tubing it poses a potential risk to the patient.
|