Customer reported that their most recent correlation yielded higher myoglobin results on the triage system than the reference lab.2 comparisons were provided (customer unable to provide all results): triage myo=108, lab=46 , triage myo=262, lab= 94.Customer stated samples were collected in edta tubes.Whole blood was first tested on the triage system and then the samples were spun down to plasma and tested at the lab.Customer stated no issues with patient results and that they were not questioning any patient results.Customer stated they have two triage meters on site and both meters correlated well with each other (serial numbers and results not provided) despite multiple attempts customer was unable to provide myoglobin reference ranges for the triage and the reference lab.No triage myoglobin results were used to determine patient care.
|