MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Device Problems Degraded (1153); Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
 
Event Description
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported a ring artifact on image data.The philips field service engineer (fse) confirmed the reported issue and evaluated the system.The fse determined the issue was due to a broken compensator in the a-plane collimator.The fse replaced the compensator to resolve the issue.The fse confirmed there was no impact to a patient and no report of misrepresentation and/or mistreatment as a result of this reported issue.The system is in clinical use.This issue was determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7243620
MDR Text Key99231256
Report Number1525965-2018-00266
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/22/2016
Initial Date FDA Received02/06/2018
Supplement Dates Manufacturer Received02/22/2016
Supplement Dates FDA Received11/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-