• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problems Device Inoperable (1663); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/09/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the system shut down during the middle of a surgical procedure.The procedure was not completed.Additional information has been requested.
 
Manufacturer Narrative
Additional information is provided.The customer reported that the system stopped working in the middle of the procedure.Additional information was requested but not obtained.The company service representative examined the system and confirmed the reported event.The company service representative reseated cables and reinstalled the software.Unrelated to the event, the system required parts to be replaced.The company service representative returned and replaced the dvd-rom drive and the u/s controller printed circuit board (pcb).The system was then tested and met all product specifications.The system was manufactured on january 28, 2009.Based on qa assessment, the product met specifications at the time of release.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7246454
MDR Text Key99328333
Report Number2028159-2018-00247
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Other Device ID Number8065750833
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2018
Initial Date FDA Received02/06/2018
Supplement Dates Manufacturer Received04/09/2018
Supplement Dates FDA Received04/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2009
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-