BOSTON SCIENTIFIC - MARLBOROUGH OBTRYX II SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
|
Back to Search Results |
|
Model Number M0068505110 |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/19/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A visual examination of the returned obtryx ii system revealed that the device was returned in its original package.The tray has witness marks around the entire perimeter confirming that the tvek lid was sealed to the tray when it left the manufacturing site. a review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Based on all gathered information and investigation results, the complaint was caused by handling of the device or portion of the device without direct patient contact either during unpacking, preparation, or shipping; at the end of the procedure; or when packing for return.Therefore, the most probable root cause classification is handling damage.
|
|
Event Description
|
It was reported to boston scientific corporation that an obtryx ii system was about to be used in a transobturator sling procedure on (b)(6) 2017.According to the complainant, during preparation when the package was opened, the seal of the sling was already open.The procedure was completed with another obtryx ii system.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|