• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN AUTOSONIX; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN AUTOSONIX; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 012001
Device Problems Failure to Cut (2587); Noise, Audible (3273)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2017
Event Type  malfunction  
Event Description
According to the reporter, during laparoscopic living donor nephrectomy procedure, a noise was heard and the device did not cut.There was no injury caused to the patient and no medical intervention was required.
 
Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.The instrument registered a mechanical fault.The instrument was disassembled; a crack was noticed in the proximal portion of the probe shaft.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the cracked probe shaft may occur when the probe shaft makes contact with the out tube while the system was energized.The information for use for this product states: avoid using the device as a lever, either while dissecting or while activating the instrument.Added stress to the tip may cause the probe to touch in the inner tube portion of the instrument, resulting in mechanical failure and/or rendering the instrument inoperable.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSONIX
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7248923
MDR Text Key99310576
Report Number1219930-2018-00659
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10884521058163
UDI-Public10884521058163
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K971861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number012001
Device Catalogue Number012001
Device Lot NumberN4F1472X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2018
Initial Date FDA Received02/07/2018
Supplement Dates Manufacturer Received03/15/2018
Supplement Dates FDA Received04/03/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-