On 02/01/2018 an ortho field engineer (fe) arrived at the customer site to investigate potential card misreads.During image review, fe found post centrifuge images are not leveled.Fe attempted to correct issue but still could not get the image to be corrected.Fe ordered a gripper and a card holders to b installed the next day.On 02/02/2018, fe returned to install the parts following the service docs.He optimized all required adjustments, customer ran qc successful.Repairs have returned this instrument to expected operation.The investigation determined that the most likely root cause of this event was instrument related.No incorrect or erroneous results were reported as a result of this incident.There was no risk present at the time of the incident.There was no harm to any patient.
|
The customer is contacting cts to report an event of provue camera misreading the antibody screen reactions.Customer repeated in manual gel, weak reactions are visually observed.Tsc had customer perform dat on the red cells and found they had a positive dat.Weak reactions are observable with naked eye, but provue is not flagging the cards.Tsc explained to the customer that the dat is not true aglutition.Customer wants service to check the provue.
|