• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 01/10/2018
Event Type  malfunction  
Manufacturer Narrative
The hospital biomed replaced the power controller board, anesthesia control board, and display connector board.Patient information could not be obtained after multiple attempts.Attempts were made as follows: (b)(6) 2018 phone call, (b)(6) 2018 phone call, (b)(6) 2018 phone call.
 
Event Description
The hospital reported that, during a case, the machine shut down.The clinician reportedly switched to manual mode to maintain ventilation.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison WI 53718
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 53188
8479711808
MDR Report Key7251201
MDR Text Key99344455
Report Number2112667-2018-00265
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042154
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/10/2018
Initial Date FDA Received02/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-