Event summary: the patient data files showed at least four applications were performed with catheter 2af284 / 43644-60 without any issues or system notices on the date of event.Data files also showed at least eight applications were performed with catheter 2af284 / 85287-74 without any issues or system notices on the date of event.Upon visual inspection of catheter 2af284 / 43644-60 results showed the device was intact with no apparent issues.Smart chip verification showed the catheter was used for four injections.The id of the push button luer was blocked (spec: 0.045").The lab test mapping catheter could not be successfully inserted into the catheter.Dissection showed excessive glue at the proximal end of the guide wire lumen shaft and the luer also guide wire lumen shaft was kinked on 1.3030 inches from the tip of the catheter.Therefore, the inability to insert the achieve has been reproduced.Catheter failed the performance test due to kink on guide wire lumen and excessive glue at the proximal end of the guide wire lumen shaft and the luer.In conclusion, the reported issue has been confirmed through testing but not confirmed through the data analysis.The balloon catheter failed the returned product inspection due to excess glue in push button luer and guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the balloon catheter shaft appeared to be kinked within the section inside the balloon.Additionally, difficulty was observed when attempting to remove the balloon catheter from the sheath.The catheter was replaced with resolve, and the case was completed with cryo.No patient complications have been reported as a result of this event.
|