Model Number NOT APPLICABLE
Device Problems Device Alarm System (1012); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2018
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunctions pressure dome membrane leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot f351 was conducted.There were no non-conformances.This lot met all release requirements.A review of kit lot f351 for the reported issue shows no trends.Trends were reviewed for complaint category pressure dome membrane leak.No trend was detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer called to report a pressure dome membrane leak during a patient treatment.The system pressure dome "popped off" the pressure transducer resulting in a blood leak on the pump deck.There was 250 ml whole blood processed at the time of the leak, and the customer stated no alarms sounded before or after the incident.The patient was in stable condition and not affected by the incident.The customer aborted the procedure with no blood returned back to the patient.There was no medical intervention required.The customer will not be returning any products back for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS
bedminster NJ
Manufacturer (Section G)
THERAKOS, INC
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key7251936
MDR Text Key100256761
Report Number2523595-2018-00026
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)F351(17)190901
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date09/01/2019
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberF351
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/11/2018
Initial Date FDA Received02/08/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age68 YR
Patient Weight97
-
-