Model Number LNQ11
Device Problem Under-Sensing (1661)
Patient Problem Stroke/CVA (1770)
Event Date 10/13/2017
Event Type  malfunction  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient had a cerebrovascular accident (cva) and the implantable cardiac monitor (icm) detected pause episodes due to suspected intermittent occasional undersensing of r-waves.The device remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7256139
MDR Text Key99839335
Report Number9614453-2018-00328
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169323575
UDI-Public00643169323575
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/14/2015
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2017
Initial Date FDA Received02/08/2018
Date Device Manufactured09/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1249-2016
Patient Sequence Number1
Patient Age80 YR
-
-