Model Number 2090
Device Problems Computer Software Problem (1112); Communication or Transmission Problem (2896); Operating System Becomes Nonfunctional (2996); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was unable to confirm the reported reboot, shutdown, and black display screen issues.However, there was a system error in the logs.The programmer hard drive was reconfigured, reloaded, and had the software updated.Analysis also noted that the power cord bay was broken, the power cord bay was replaced.The programmer passed all final functional an and system tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer was unable to complete an interrogation because the programmer would start then reboot, start and then shutdown, and unexpectedly displayed a black screen multiple times during the interrogation.The programmer was returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7256657
MDR Text Key99898842
Report Number2182208-2018-00198
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581570
UDI-Public00613994581570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received02/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-