Model Number 2090
Device Problems Break (1069); Operating System Becomes Nonfunctional (2996)
Patient Problem No Patient Involvement (2645)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
Analysis summary: analysis was unable to confirm nor reproduce the customer comment that the programmer booted down while in use, it passed its incoming vxi testing and no error was found in the logs.It was noted that the cord door latch tab was broken as was the left keyboard hinge.The cord bay and the keyboard hinge were both replaced.Analysis did confirm the customer comment that the handle was broken and the handle was replaced.The programmer passed all console and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer boots down while in use.It was additionally noted that the handle was broken.The programmer was returned for service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7257772
MDR Text Key99892535
Report Number2182208-2018-00264
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/06/2017
Initial Date FDA Received02/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-