ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH23
Device Problems Break (1069); Detachment Of Device Component (1104); Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p93d31.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.In addition the shaft was intact and no broke off as the customer reported.During functional testing on gen11 an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during an unknown procedure, the shaft broke off and loosened.No patient consequence reported.
 
Manufacturer Narrative
(b)(4).Additional information: the nurse stated that when they started to use the harmonic, first the surgeon noticed that the harmonic shaft was little bit loose ( not properly closed in the handle).And then the shaft was separated from the handle totally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 W ADV HEMOSTASIS
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7259543
MDR Text Key99993081
Report Number3005075853-2018-07881
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014673
UDI-Public20705036014673
Combination Product (y/n)N
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberHARH23
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/30/2018
Initial Date FDA Received02/09/2018
Supplement Dates Manufacturer Received01/30/2018
Supplement Dates FDA Received02/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N