Hill-rom technical support was set to arrange a service call for a field technician to evaluate and repair the bed.Per the hill-rom service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Head section motor: inspect the actuator assembly.Make sure the pins and retaining clips are intact and not missing.Fully raise and lower the head section.Make sure there is no friction or abnormal noises and that no audible overload indication can be heard during the movement.Repair as necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account contacted hill-rom technical support and stated that they no longer want the bed repaired as the patient does not have a need for it at this time.The account requested the repair be cancelled.Based on this information, no further action is required.
|