Model Number U1170G0000307
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2018
Event Type  malfunction  
Manufacturer Narrative
Hill-rom technical support was set to arrange a service call for a field technician to evaluate and repair the bed.Per the hill-rom service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Head section motor: inspect the actuator assembly.Make sure the pins and retaining clips are intact and not missing.Fully raise and lower the head section.Make sure there is no friction or abnormal noises and that no audible overload indication can be heard during the movement.Repair as necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account contacted hill-rom technical support and stated that they no longer want the bed repaired as the patient does not have a need for it at this time.The account requested the repair be cancelled.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head and foot sections experienced self run.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREASSIST ES
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key7259667
MDR Text Key100002203
Report Number3006697241-2018-00012
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Patient
Type of Report Initial
Report Date 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberU1170G0000307
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/10/2018
Initial Date FDA Received02/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-