MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARHD36
Device Problem Loss of Power (1475)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.The reported complaint was that the device stopped working.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.The activation issues are related with the blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿blade error detected¿ or ¿relax pressure on blade¿ or ¿remove instrument from patient¿ followed by a ¿replace instrument¿ screen displayed on the generator.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during a laparoscopic hysterectomy procedure the device stopped working.No additional information is available.The procedure was completed using a like device of the same product code.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC HD 1000I SHEARS 36CM SHAFT
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7260196
MDR Text Key99992199
Report Number3005075853-2018-07891
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036015052
UDI-Public20705036015052
Combination Product (y/n)N
PMA/PMN Number
K160752
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberHARHD36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2018
Initial Date Manufacturer Received 01/31/2018
Initial Date FDA Received02/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-