It was reported that the patient was experiencing a tingling sensation at the electrodes that traveled from her ear to her head.Diagnostics were performed after the tingling started, and high impedance was present.The patient used the magnet to disable the device, and she planned on asking her physician to program it off.No surgical intervention has occurred to date.No further relevant information has been received to date.
|
Analysis on the lead was approved.Note that a small portion of the lead assembly including the electrodes was not returned for analysis, and therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the returned portion of the lead, one of the quadfilar coils appeared to be broken close to the end of the abraded open / cut outer silicone tubing.Scanning electron microscopy was performed and identified the area on three of the broken coil strands as being mechanically damaged, which prevented identification of the coil fracture type with fine pitting and residual material on one, fine pitting on another, and no pitting and residual material on the third one.The area on the remaining broken coil strand was identified as having evidence of a stress induced fracture (fatigue appearance) with mechanical damage and fine pitting.Pitting was observed on the coil surface.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.The abraded openings found on the outer and inner silicone tubes, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer and inner silicone tubes.With the exception of the observed discontinuity and abraded openings observed on the outer and inner silicone tubes, the condition of the returned lead portions was consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, and no other discontinuities were identified.Analysis on the generator was also approved.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrated that accurate resistance measurements were obtained in all instances.Proper functionality of the pulse generator in its ability to provide appropriate programmed output currents was successfully verified.In addition, the septum was not cored, thus eliminating the possibility of a potential unintended electrical current path through body fluids.The device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.No further relevant information has been received to date.
|