• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the laser was not working.Additional information has been requested.
 
Manufacturer Narrative
Additional information is provided.The customer spoke to the company service representative to assist with troubleshooting.The customer found that the emergency stop switch was pushed in by accident.Once it was pulled back into its retracted position, the laser was working normally.The customer was able to troubleshoot and resolved the problem reported.The customer did not request service.The system was manufactured on april 20, 2010.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the user accidentally pressing the emergency stop button.Based on the information provided, the emergency stop switch functioned as intended and no problem was found with the system.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7265317
MDR Text Key100045352
Report Number2028159-2018-00290
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/19/2020
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number8065751150
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/20/2018
Initial Date FDA Received02/12/2018
Supplement Dates Manufacturer Received04/02/2018
Supplement Dates FDA Received04/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-