• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5_12.1
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/27/2017
Event Type  malfunction  
Manufacturer Narrative
This product is not marketed in the us.Lens work order search: no similar complaint type events reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 12.1 mm vicm5_12.1, -9.5 diopter, implantable collamer lens into the patient's right eye (od) on (b)(6) 2017.After implant the surgeon states he found a "tiny black-colored foreign substance on the lens surface." reportedly, the surgeon attempted to remove the substance but was unsuccessful.The surgeon removed the lens intra-operatively.On the same date, in a second surgery, an alternate lens was successfully implanted.
 
Manufacturer Narrative
Returned product evaluation found lens returned dry in a micro centrifuge vial with clear residue on lens.Visual inspection found a piece of the haptic missing and residue on lens.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
8002927902
MDR Report Key7267545
MDR Text Key100259888
Report Number2023826-2018-00216
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/30/2019
Device Model NumberVICM5_12.1
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/19/2018
Initial Date FDA Received02/13/2018
Supplement Dates Manufacturer Received03/12/2018
Supplement Dates FDA Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
-
-