MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A doctor report the product occluded and did not work during a procedure.The product replaced and procedure completed with no patient harm.A product sample has been requested; however, it has not been received for evaluation at the manufacturing site.
 
Manufacturer Narrative
This is the seventh complaint reported for this finished goods lot; however the sixth for this issue.Review of the device history record indicates the order was built to specification.The returned cassette was visually and functionally tested and passed testing.No contributing factors could be identified that could cause the customer's reported issue.The root cause of the customer's complaint could not be established; the returned sample met specifications.After a thorough investigation of this complaint, it has been determined that no action will be taken at this time as the sample met specifications.Quality assurance will continue to monitor customer complaints via the complaint review meetings, and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7268346
MDR Text Key100132742
Report Number2028159-2018-00295
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Other Device ID Number380657517633
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2018
Initial Date FDA Received02/13/2018
Supplement Dates Manufacturer Received05/10/2018
Supplement Dates FDA Received05/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-