On (b)(6) 2017, a gore® acuseal vascular graft was implanted into the patient¿s left underarm in order to perform dialysis treatments up to three times per week.It was reported to gore that on (b)(6) 2018, the patient was diagnosed with an occlusion of the gore® acuseal vascular graft.Subsequently, a thrombectomy was performed with a cleaner device from argon medical devices (6f ¿ 65cm x 9mm) followed by a percutaneous transluminal angioplasty with a 6mm x 40mm balloon.It was stated that the intervention achieved good results.One day after the reintervention procedure a reocclusion of the gore® acuseal vascular graft was noticed.Therefore an acute open surgery was performed where a small portion of the graft was explanted.A visual inspection of the explanted graft portion indicated a separation of the ptfe layers of the graft.It was reported that it is assumed that this separation caused the medical device occlusion.Consequently, the gore® acuseal vascular graft was disconnected but still remains in the patient next to a new implanted gore® acuseal vascular graft.It was stated that the patient was doing well following the procedure.
|
The review of the manufacturing records verified that this lot met all pre-release specifications.The device remains implanted, just one little portion of the graft was explanted and visually inspected by the hospital.As this portion was discarded afterwards, an engineering evaluation could not be performed.
|