• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED; MQR STENT, COLONIC METALLIC EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED; MQR STENT, COLONIC METALLIC EXPANDABLE Back to Search Results
Model Number G48028
Device Problem Delivery System Failure (2905)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) # k163468.(b)(4).Problem statement: at preparation stage, operator want to confirm there is no issue of releasing the stent so that he checked the stent release before use while found out the stent cannot be released after pull the trigger.Operator confirmed that the red marker worked properly with no issue.Device evaluation: the evo-25-30-8-c device of lot number c1273067 was returned to cook (b)(4) and evaluated on the 17 jan 2018.Lab evaluation: upon evaluation of the returned device, the lockwire was still in place.Upon actuation, no deployment or retraction was possible.The red shuttle deployment marker was at the back half of the handle.There was no stent exposure on return.The device was dismantled during the lab, and the flexor was noted to be broken at the shuttle cap.The stent was manually deployed during the lab and there were no issues noted with the stent.Customer complaint confirmed as failure was verified in laboratory.The flexor was seen to be broken at the shuttle cap.Root cause: a possible cause for the issue occurring may be due to delamination of the ptfe liner of the outer sheath document review: prior to distribution all evo-25-30-8-c devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at (b)(4).A review of the qc records did not reveal any issues which could have contributed to this complaint issue.Ifu review: as per the instructions for use, which accompanies this device instructs the user to "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".On review of the information provided, there is no evidence to suggest that the user did not follow the instructions for use.Summary: customer complaint confirmed as failure was verified in laboratory.The flexor was seen to be broken at the shuttle cap.From the information provided, there have been no adverse effects to the patient as a result of this occurrence.This event occurred prior to patient contact.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
Initial mdr is being submitted based on the device malfunction precedence: ¿flexor kinked/stretched/broke/compressed".At the preparation stage, the operator wanted to confirm there was no issue releasing the stent so he checked that the stent would release before use.He found the stent cannot be released after he pulled the trigger.Operator confirmed that the red marker worked properly with no issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED
Type of Device
MQR STENT, COLONIC METALLIC EXPANDABLE
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer Contact
michael galvin
o halloran road
national technology park
limerick 
061334440
MDR Report Key7269993
MDR Text Key100432821
Report Number3001845648-2018-00063
Device Sequence Number1
Product Code MQR
UDI-Device Identifier10827002480282
UDI-Public(01)10827002480282(17)180913(10)C1273067
Combination Product (y/n)N
Reporter Country CodeCN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG48028
Device Catalogue NumberEVO-25-30-8-C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/14/2018
Event Location Hospital
Initial Date Manufacturer Received 01/17/2018
Initial Date FDA Received02/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-