Model Number H7495551110
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 01/23/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that guidewire separation occurred.The target lesion was located in the distal left anterior descending artery (lad).While trying to advance the comet pressure guidewire into a very tortuous saphenous vein graft (svg) in an attempt to analyze the distal lad, the physician realized that the wire could not be advanced any farther.The physician then tried to remove the wire and ¿felt no resistance.¿ upon removal, physician recognized that the wire was broken about 30cm from the distal tip.Physician was able to utilize a gooseneck snare and successfully removed the wire tip from the body.The physician then used a non-bsc wire and did successful intervention to the lad lesion.Slight delay in procedure but patient was treated successfully without harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7270469
MDR Text Key100062917
Report Number2134265-2018-01015
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2019
Device Model NumberH7495551110
Device Lot Number0020886814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2018
Initial Date FDA Received02/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-