Investigation summary: the returned device was visually inspected and it was found in normal conditions.Per the event, the catheter was tested for electrical performance and it was found within specifications.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.The customer complaint cannot be confirmed. concomitant product: non biosense webster, inc.- biotronik pacing stimulator.Manufacturer's ref.No:(b)(4).
|
It was reported that a patient underwent a procedure with a navistar¿ electrophysiology catheter.During the procedure, there was noise on the navistar¿ electrophysiology catheter.It was also reported that when pacing, the catheter paced through all channels.The catheter was replaced and the issues resolved.There was no patient consequence reported.Additional information was received stating that the pacing leads were connected to the primary pacing port.There was unwanted pacing being delivered as the system paced through all the channels at the same time.No ablation was performed before replacing the navistar¿ electrophysiology catheter.The biotronik pacing stimulator was being used during the procedure.The noise issue was assessed as not reportable.Ecg noise was only indicated on the navistar¿ electrophysiology catheter.Therefore,the risk to the patient was low.The pacing issue was assessed as a reportable malfunction.
|