• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE 64; COMPUTED TOMOGRAPHY X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE 64; COMPUTED TOMOGRAPHY X-RAY Back to Search Results
Model Number 728231
Device Problems Disconnection (1171); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported cables were not reconnected following a previous service visit and the system auto-voice feature was not working as a result.However, philips remote support engineer (rse) confirmed that the microphone line in and line out cable was disconnected.As a result, the operator would not be able to hear the patient.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.Based on the further information, this has been determined to be a reportable event.
 
Manufacturer Narrative
On (b)(6) 2017, the customer reported that the auto-voice feature was not working, nor could the operator hear the patient from the scan room, after a recent service event (pr7502237) to replace a graphics card.The customer confirmed with the philips helpdesk (when this issue was reported) that there was no patient impact and no harm as a result of this event.The philips remote service engineer (rse) instructed the customer to move the host computer to check the audio cable connection status; the audio cables were not connected.The rse instructed the customer how to re-connect the line out audio cable affecting auto-voice and the line in cable from the gantry microphone to the host to resolve the issue.The system is operational and in clinical use.This event has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE 64
Type of Device
COMPUTED TOMOGRAPHY X-RAY
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
Manufacturer Contact
derrick massuri
595 miner rd
cleveland, OH 44143
4404833000
MDR Report Key7272613
MDR Text Key100434013
Report Number1525965-2018-00349
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K033326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728231
Device Catalogue NumberNCTB423
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/28/2017
Initial Date FDA Received02/15/2018
Supplement Dates Manufacturer Received12/28/2017
Supplement Dates FDA Received08/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-