• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080617
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2018
Event Type  malfunction  
Manufacturer Narrative
Mfg date: the device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Manufacturer Narrative
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: failed insulation.The probable root causes could be user misuse, excessive force, normal wear, or improper sterilization methods.Although the product passed the insulscan, the failure identified in the investigation is consistent with the complaint record because there were dents and scratches on the insulation.Manufacture date is not known.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM PEEK MONOPOLAR HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
roomi banerjee dua
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7274073
MDR Text Key100439589
Report Number0002936485-2018-00131
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080617
Device Lot Number¿0847912D¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/18/2018
Initial Date FDA Received02/15/2018
Supplement Dates Manufacturer Received01/18/2018
Supplement Dates FDA Received05/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-