Model Number EDNS-9001
Device Problems No Display/Image (1183); Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/19/2018
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the prefense monitor's screen went blank and emitted no sound.They were provided with the part number for a new lcd display to resolve the issue.No patient harm was reported.
 
Event Description
The customer reported that the prefense monitor's screen went blank and emitted no sound.
 
Manufacturer Narrative
H10: additional narrative: the customer reported that their monitor goes blank with no sound.Service requested: the customer requested an exchange for the defective unit.Service provided: the customer was provided with part number for the defective unit.Investigation result: the prefense (a/edns-9001; sn: (b)(6) was placed into service on 10/31/2011, which is 7 years prior to the reported issue.A/edns-9001 service manual revision i provides guidance for troubleshooting, servicing, maintaining or repairing the prefense edns-9001.The service manual involves troubleshooting of the display and advises customer on proper action to take.Previously reported issues with this unit indicates that the user error/environmental factor was prevalent with this customer.The root cause of the monitor going blank was unable to determine.The cause of the issue was determined to be failure of the display unit.The device was in use with a patient and there was no reported patient harm.Corrected information: g4.Date received by manufacturer: should be 01/19/2018 not 02/16/2018 as listed on mdr initial report.Additional information: b4.Date of this report.D4.Unique identifier (udi) number: pre-udi requirement.E1.Name and address.E2.Health professional.E3.Occupation.F3-5.Contact for user facility/importer.F6.Date user facility/importer became aware of the event.F7.Type of report.F9.Approximate age of device f11.Date report sent to fda f12.Location where event occurred f13.Date report sent to manufacturer g4.Date received by manufacturer g7.Type of report h2.If follow-up, what type? additional information.Correction.H3.Device evaluated by manufacturer? h6.Event problem and evaluation codes.H10.Additional manufacturer narrative.
 
Event Description
The customer reported that the prefense monitor's screen went blank and emitted no sound.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRAL MONITORING SYSTEM
Type of Device
CENTRAL MONITORING STATION
Manufacturer (Section D)
NKUS LAB
5 jenner, suite 180
attn: shama mooman
irvine CA 92618
MDR Report Key7277947
MDR Text Key100522791
Report Number2032233-2018-00045
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
K102106
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEDNS-9001
Device Catalogue NumberEDNS-9001
Device Lot NumberNOT APPLICABLE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/24/2019
Distributor Facility Aware Date07/24/2019
Device Age81 MO
Date Report to Manufacturer07/24/2019
Initial Date Manufacturer Received 02/16/2018
Initial Date FDA Received02/16/2018
Supplement Dates Manufacturer Received07/24/2019
Supplement Dates FDA Received07/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNOT APPLICABLE
Patient Sequence Number1
-
-