• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON VISION CARE, INC. ¿ US ACUVUE OASYS1 DAY WITH HYDRALUXE TECHNOLOGY; LENSES, SOFT CONTACT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON VISION CARE, INC. ¿ US ACUVUE OASYS1 DAY WITH HYDRALUXE TECHNOLOGY; LENSES, SOFT CONTACT Back to Search Results
Catalog Number TSP
Device Problem No Apparent Adverse Event (3189)
Patient Problems Irritation (1941); Pain (1994); Red Eye(s) (2038)
Event Date 01/02/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).(b)(6).
 
Event Description
On (b)(6) 2018 an eye care provider (ecp) in (b)(6) reported a patient (pt) had ¿used a daily lens due to the lack of usual monthly lens¿ shortly after placing the lens the pt experienced a stinging sensation which escalated to sharp pain¿.The pt sought medical assistance with an ecp ¿who could see damage to conjunctiva and scraping of cornea.Lens rest and citodon¿.New appointment on (b)(6) 2018, ecp ¿suspects toxicity of cornea¿.Prescribed opnol and was given and appointment for follow-up.The suspect product was reported as acuvue oasys 1 day brand contact lenses.On 23jan2018 a call was placed to the pts optometrist and additional medical information was provided: both eyes were affected.Pt told the optometrist no solution had been used in combination with the lenses prior to insertion.Additional information was requested.On 29jan2018 additional information was received from the affiliate in (b)(6): the patient visits the hospital on (b)(6) 2018.¿the pt suffers from blurred vision since the day before.The pt inserted the daily disposables for the first time, but the blurriness still remained, the pt took out the lenses and immediately felt pain in the left eye and is not able to open the eye¿.Pt experienced photophobia.Pt reported no trauma involved.When removing the contact lens in the left eye, irritation and erosion were visible.Irritation were found in the right eye as well, but no erosion.Pt had a follow-up visit a few days later: left eye is ok, still irritation in the right eye.Finding according to the ecp is that the right eye has no ¿pus¿, but moderate bulbar hyperemia.Small clouding in corneal stroma.Oedema.Left eye is now good with clear, transparent cornea.Doctor suspect toxicity caused by the contact lenses.Treatment is opnol x3 (cortison)¿.No additional information was provided.No additional information has been received.This report is for the pts left eye event.The event for the pts right eye will be submitted in a separate report.It is unknown if the suspect product is available for return.A lot history review was performed and revealed the following: the batch record did not show any abnormalities in monomer and solution testing.All parameters tested were within specification.All sterilization requirements were successfully completed.Lot 5520870101 was produced under normal conditions.If additional information is received it will be reported within 30 days of receipt.Serious reportable event trends are reviewed quarterly in franchise management review meetings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUVUE OASYS1 DAY WITH HYDRALUXE TECHNOLOGY
Type of Device
LENSES, SOFT CONTACT
Manufacturer (Section D)
JOHNSON & JOHNSON VISION CARE, INC. ¿ US
7500 centurion parkway
jacksonville FL
Manufacturer Contact
rose harrell
7500 centurion parkway
jacksonville, FL 32256
9044433364
MDR Report Key7278116
MDR Text Key100310735
Report Number1057985-2018-00018
Device Sequence Number1
Product Code LPL
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K042275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/01/2020
Device Catalogue NumberTSP
Device Lot Number5520870101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2018
Initial Date FDA Received02/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2015
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-