• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE 16 AIR; COMPUTED TOMOGRAPHY X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE 16 AIR; COMPUTED TOMOGRAPHY X-RAY Back to Search Results
Model Number 728246
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported the speaker from the gantry to the console was not working.Philips service verified the operator was not able to hear the patient at all times.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
 
Manufacturer Narrative
The customer reported on (b)(6) 2018 that the operator could not hear the patient in the scan room.The customer confirmed with the philips helpdesk (when this issue was reported) that there was no patient impact and no harm as a result of this event.The philips field service engineer (fse) arrived at the customer site on (b)(6) 2018 to evaluate the ct system.The fse confirmed the customer¿s allegation that no sound could be heard from the ct gantry to the ct box.The fse determined that the microphone board failed; the volume on the ct box had no issue.The fse replaced the breathing light assembly, which includes a microphone module along with the right front microphone to gantry audio board cable and left front microphone to gantry audio cable to correct and resolve the issue.The system is operational and in clinical use.This event has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE 16 AIR
Type of Device
COMPUTED TOMOGRAPHY X-RAY
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7278171
MDR Text Key100520048
Report Number1525965-2018-00352
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K012009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728246
Device Catalogue NumberNCTA399
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/24/2018
Initial Date FDA Received02/16/2018
Supplement Dates Manufacturer Received01/24/2018
Supplement Dates FDA Received08/22/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-