• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL MIDWEST E PLUS 1:5 HIGHSPEED ATTACHMENT; HANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL MIDWEST E PLUS 1:5 HIGHSPEED ATTACHMENT; HANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL Back to Search Results
Catalog Number 875205
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Burn, Thermal (2530)
Event Type  malfunction  
Manufacturer Narrative
Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event, a doctor reported that a midwest e plus1:5 attachment overheated and burned a patient's lip.There was no intervention required.
 
Manufacturer Narrative
Evaluation found bur bearing damaged/ bearing debris in head from failed bur bearing.Cap also has rotor mark on cap.Replaced turbine and new cap.Cleaned water port, replaced water tubes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E PLUS 1:5 HIGHSPEED ATTACHMENT
Type of Device
HANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7278687
MDR Text Key100525679
Report Number1419322-2018-00020
Device Sequence Number1
Product Code EFA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K965238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number875205
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/18/2018
Initial Date FDA Received02/16/2018
Supplement Dates Manufacturer Received04/27/2018
Supplement Dates FDA Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-