MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Failure to Deliver Energy (1211); Ambient Noise Problem (2877)
Patient Problems Nausea (1970); Therapeutic Effects, Unexpected (2099)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins) for gastric stimulation.It was reported that the patient had surgery and the surgery was not related to the device/therapy; however, the patient had an ultrasound done during the surgery and since the surgery, the patient was nauseous so they wanted to check the device to make sure it was still working.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a consumer.It was reported that the patient went to see the implanting physician and the device was off; it was turned back on and reprogrammed, so the device was not damaged.It was noted that the only way it could have been turned off was by the ultrasound that was used during the surgery.It was reported that after the surgery, the patient was very sick.The illness resolved by turning on and reprogramming, the patient was doing well now, and the device really helped with their nausea.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7280912
MDR Text Key100530449
Report Number3004209178-2018-03572
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2019
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/16/2018
Initial Date FDA Received02/19/2018
Supplement Dates Manufacturer Received03/01/2018
Supplement Dates FDA Received03/19/2018
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
-
-