MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500 Back to Search Results
Model Number 500AHCT
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the monitor to have no viewable liquid crystal display (lcd) display back ground.The lcd invertor printed circuit board assembly (pcba) revealed a hi voltage coil was opened or shorted, producing heat which melted the surrounding cover on the pcba.The pcba no longer produces the required alternating current (a/c) voltage to drive or operate the lcd display properly.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the display on the unit was dark.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7283943
MDR Text Key100666067
Report Number1828100-2018-00084
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/20/2018
Initial Date FDA Received02/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-