Catalog Number 2C4005
Device Problem Device Packaging Compromised (2916)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the sterile packaging of three (3) y-type tur irrigation sets was open.This event was discovered before patient use during the kitting process.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Two devices were received for evaluation.During visual examination, the pouch was observed open; therefore, the peel and seal width test was performed to evaluate the quality of the seal.The results were satisfactory due to the sealed is uniformed across the sealed area.Also, it was observed that the packaging material is correctly aligned.The reported condition was verified; however, the cause of the condition could not be determined.There are controls in place to detect and prevent an opening packaging in the manufacturing.The third device was not returned for evaluation; therefore, a device analysis could not be completed.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUR IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7284494
MDR Text Key100526420
Report Number1416980-2018-00912
Device Sequence Number1
Product Code LJH
UDI-Device Identifier00085412000978
UDI-Public(01)00085412000978
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/18/2022
Device Catalogue Number2C4005
Device Lot NumberR17G18057
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/25/2018
Initial Date FDA Received02/20/2018
Supplement Dates Manufacturer Received03/12/2018
Supplement Dates FDA Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-