(b)(4).Method: the complaint mr810 humidifier was received at our fisher & paykel healthcare (fph) regional office in (b)(4), where it was inspected and performance tested by a trained fph technician.Our investigation is therefore based on the investigation conducted by the (b)(4) office and the information provided by the hospital.Result: the performance testing of the subject mr810 humidifier revealed that there were no faults with the humidifier.Conclusion: the mr810 has not caused or contributed to the event.The intersurgical mg5804855 circuit was used with the mr810 humidifier, and is not an approved fph product.The hospital confirmed that an unheated extension was used with the circuit.This would have likely been the reason for the excessive condensation reported by the customer.Our user instructions that accompany the mr810 respiratory humidifier states: the use of breathing circuits, chambers or other accessories which are not approved by fisher & paykel healthcare may impair performance or compromise safety.Condensate in the humidification system is an expected side effect of heated pass-over humidification systems in many conditions, and may vary between light misting to water droplets that form on the wall of cool breathing circuit tubing.The amount of condensate in the ventilation system is influenced by a number of multiple set up and environmental factors, including use of third party breathing circuits not approved by fph.The mr810 is equipped with several safety features that limit the risk of thermal injury.In the event of a failure the unit will shut down in a safe state and warn the user that a fault has occurred.It does this by monitoring the correct operation of its critical circuits.If a fault is detected it will try to report this by alarming visually.
|