Model Number 6379
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/29/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The fsr found the drive motor speed to be at 3740 revolutions per minute (rpm) when the maximum speed should be 3600 +/- 100 rpm.He adjusted the maximum speed to be within specification.The unit operated to the manufacturer¿s specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the drive motor speed went beyond the maximum speed specification.There was no patient involvement.
 
Manufacturer Narrative
Please disregard the initial emdr submission for this complaint.It has been determined that this complaint is not valid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7285113
MDR Text Key100671842
Report Number1828100-2018-00085
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number6379
Device Catalogue Number6379
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/29/2018
Initial Date FDA Received02/21/2018
Supplement Dates Manufacturer Received03/16/2018
Supplement Dates FDA Received03/16/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-