Model Number 6379 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The fsr found the drive motor speed to be at 3740 revolutions per minute (rpm) when the maximum speed should be 3600 +/- 100 rpm.He adjusted the maximum speed to be within specification.The unit operated to the manufacturer¿s specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the drive motor speed went beyond the maximum speed specification.There was no patient involvement.
|
|
Manufacturer Narrative
|
Please disregard the initial emdr submission for this complaint.It has been determined that this complaint is not valid.
|
|