• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 4.5MM TAP (CANNULATED); ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 4.5MM TAP (CANNULATED); ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734298
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/26/2018
Event Type  malfunction  
Manufacturer Narrative
Patient identifier field not sufficient to hold all digits, should read: (b)(6).The suspect tap has been received by the manufacturer for evaluation.The reported issue was confirmed as the tip of the instrument has been bent and broken off.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A site representative reported that during spinal fusion procedure, the tip of the tap was bent.It was reported that the issue occurred when the surgeon was malleting pelvic bone on the patient.The procedure was completed with the use of navigation.There was no delay to procedure.No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM TAP (CANNULATED)
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7285712
MDR Text Key100669288
Report Number1723170-2018-00786
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734298
Device Lot Number160816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2018
Initial Date FDA Received02/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age28 YR
Patient Weight37
-
-