Model Number DM3500
Device Problems Migration or Expulsion of Device (1395); Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2018
Event Type  malfunction  
Event Description
It was reported that while implanting the implantable cardiac monitor while suturing, the device header migrated out of the pocket and sensing was lost.The device was repositioned and the sensing could not be restored.A second pocket was made and no sensing could be established with the device in it either.The device was not used and a new device was implanted.The patient was fine throughout the procedure and will continue to be monitored.
 
Manufacturer Narrative
The reported field event of loss of sensing was not confirmed in the laboratory.The device was tested and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7288743
MDR Text Key100733403
Report Number2017865-2018-02929
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2019
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000051045
Other Device ID Number05415067027320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/01/2018
Initial Date FDA Received02/22/2018
Supplement Dates Manufacturer Received02/28/2018
Supplement Dates FDA Received03/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1