MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; RSP STANDARD HUMERAL SOCKET INSERT, 32MM, HXE-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; RSP STANDARD HUMERAL SOCKET INSERT, 32MM, HXE-PLUS Back to Search Results
Model Number 509-00-032
Device Problems Unstable (1667); Device Operates Differently Than Expected (2913); Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); No Code Available (3191)
Event Date 02/12/2018
Event Type  Injury  
Event Description
Revision surgery - the patient had instability and shoulder was dislocating.
 
Manufacturer Narrative
The reason for this revision surgery was due to the joint instability and dislocation.The previous surgery and the revision detailed in this investigation occurred over 1 year and 2 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the joint instability and dislocation.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: loose joint from inadequate soft tissue support, degenerative bone, patient bone deterioration, excessive range of motion, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
RSP STANDARD HUMERAL SOCKET INSERT, 32MM, HXE-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key7292773
MDR Text Key100815478
Report Number1644408-2018-00162
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912144537
UDI-Public(01)00888912144537
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/11/2021
Device Model Number509-00-032
Device Catalogue Number509-00-032
Device Lot Number377P1179
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2018
Initial Date FDA Received02/23/2018
Supplement Dates Manufacturer Received03/01/2018
Supplement Dates FDA Received03/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
508-32-101, LOT 862C2254
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-