|
Model Number AJR7112-GB |
Device Problems
Misassembled (1398); Failure to Align (2522)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Additional information will be provided upon investigation conclusion.
|
|
Event Description
|
On 2018-jan-24 arjohuntleigh was notified by customer facility about event with involvement of malibu bathtub (serial number: (b)(4)), which occurred on (b)(6) 2018.The caregivers were transferring a patient with the bath chair when they noticed that the chair was sitting at an angle.They moved the patient safely and discovered that the chair safety clip appeared to be reversed.The involved patient did not sustain injuries.When on 2018-jan-25 the arjohuntleig representative attended customer facility, no issues were found.The customer facility representative confirmed that they had refitted the safety clips on bath.
|
|
Manufacturer Narrative
|
On 2018-jan-24 arjohuntleigh was notified by customer facility about event with involvement of malibu bathtub (serial number: see0346259), which occurred on 2018-jan-12.The caregivers were transferring a patient with the use of the bath chair when they noticed that the chair was set at an angle.They positioned the patient safely back into the bath? and discovered that the chair safety clip (locking fitting) was installed conversely.The involved patient did not sustain injuries and no treatment was required.When on 2018-jan-25 the arjohuntleigh representative attended customer facility, no issues were found during the device evaluation.The customer facility representative confirmed that they had refitted the safety clip on bath.The further evaluation of the device revealed that it was in good condition and no other malfunction was detected.The review of similar reportable events with the involvement of the malibu in last 5 years revealed that number of events related to situation where safety clip was not properly locked is very low.The involved malibu bath was under the arjohuntleigh maintenance contract and last service was performed on 2017-dec-04, when the arjohuntleigh representative replaced leaking shower hose and refitted loose collar between outer tube and lift arm.The device was also serviced on 2017-jun-06 (replacement of middle pipe and lift arm) and 2017-may-09 (annual maintenance).Following the reports of services performed by the arjohuntleigh representative no abnormality related to the safety catch was detected.This was the first time the customer facility notified the arjohuntleigh representative about this kind of malfunction.It was also confirmed that safety catch/clip was never replaced.Please note that malibu operating and product care instructions [04.Au.02-3gb dated on march 2003] which was delivered to customer with the claimed device provides information and warnings related to safe usage of the bathtub including locking and securing the lift chair: "warning! before using the transfer chair, make sure it is safely attached to the lift arm and that the locking mechanism is thoroughly locked." (p.4) "warning! before using the transfer chair always make sure that it is safely attached to the lift arm." (p.15) the opci provides also instructions regarding correct connecting and disconnecting the transfer chair including pictures showing the safety catch locking as well (p.11-12).The instructions show the proper positioning of chair or application of safety catch and should be confronted with the condition of device to prevent from event occurrence."make sure the hooks are securely attached to the lift arm and that safety catch is in place." according to the preventive maintenance schedule provided within opci the qualified personnel during service of the device should: "check/replaced safety catches on lift arm and chassis." (p.21) on 2017-jun-06 the arjohuntleigh representative replaced middle pipe and lift arm on which the safety clip for bath chair is fitted.However, this service was performed over 7 months before the issue was first time noticed by the customer.The involved device was under the arjohuntleigh service contract and its preventive maintenance was performed in 2017-dec-04.According to this information no malfunction of the safety catch was noted during the maintenance and recorded in service report.It should be underlined that opci informs user that device alterations may cause potential hazard for the resident: "unauthorized modifications on any arjohuntleigh equipment may affect its safety." (p.3) please note that when arjohuntleigh representative visited the customer facility to evaluate the device it was already rectified - safety catch was retrofitted.Based on the information provided by customer facility safety catch was reversed (turned upside down).However, as per the technical review of this malfunction it should not cause the effect, which was described.The design of part would make it improbable to assemble it with lock fitting turned 180 degrees.Based on the collected information the involved device failed to meet the manufacturer's specification as the safety clip was wrongly fitted.At the time of event the device was used for treatment of the patient (customer of the facility).Please note that no adverse event occurred in this case.This complaint was decided to be reported to the competent authorities in abundance of caution, due to potential of serious injury if the event were to reoccur and if the malfunction was not detected.Also, because of the limited information available justifying why safety catch were retrofitted by customer facility before the bathtub was made available for arjohuntleigh representative evaluation.Please note that fields f.3-5 were corrected due to change of contact person and fields g.1-2 were modified because of e-mail address change.
|
|
Manufacturer Narrative
|
Information regarding visit at the customer facility which took place on 2018-jan-25 were found to be inaccurate and required clarification.This is a follow-up to the report (reported under manufacturer report number: 3007420694-2018-00046) to present revised information collected and investigation conclusion: on 2018-jan-24 arjohuntleigh was notified by the customer facility about an issue found on malibu bathtub (serial number: (b)(4)), which was detected on 2018-jan-12.The caregivers were transferring a patient with the bath chair when they noticed that the chair was at an angle.After they safely took the patient from chair it was discovered that the chair locking fitting element appeared to be mounted conversely.This however did not affect the chair stability.The resident transfer was completed safely without any issues.No injury nor adverse event occurred.On 2018-jan-25 the arjohuntleigh representative attended the customer facility and during the device inspection confirmed that locking fitting was not attached correctly and repositioned this element.The further evaluation of the bathtub revealed good technical device condition and lack of any other deviation.The review of reportable events with the involvement of the malibu in last 5 years revealed that this was a first time arjohuntleigh was notified about this kind of complaint.Therefore we deem this was a single occurrence issue.The malibu/sovereign bathtub is equipped with detachable chair and lift arm for transfer of the patient.The lift arm is attached to the actuator by three countersunk bolts.The bath chair is equipped with two hooks on which the chair hangs when is mounted on the lifting arm.Moreover, the lifting arm has a guide pin which while the chair is being attached to the lift arm is inserted into the chair and provides additional support.According to the bathtub evaluation results, the device was in good condition and all these described elements of lifting arm were present on the device.It was confirmed that the construction of bathtub lifting arm provided proper stability of the chair.Based on this information the likelihood of occurrence of serious deterioration in state of health was found negligible and it do not suggest this issue caused by itself any immediate risk for a patient.This complaint was initially reported to the competent authority in abundance of caution, based on initial allegation suggesting instability of the device chair.Please note that malibu operating and product care instructions [opci 04.Au.02-3gb dated on march 2003] which was delivered to customer with the claimed device provides information and warnings related to safe usage of the bathtub including locking and securing the lift chair."warning! before using the transfer chair, make sure it is safely attached to the lift arm and that the locking mechanism is thoroughly locked." (p.4) "warning! before using the transfer chair always make sure that it is safely attached to the lift arm." (p.15).The opci provides also instructions regarding correct connecting and disconnecting the transfer chair including figures showing how to attach the locking fitting properly."make sure the hooks are securely attached to the lift arm and that safety catch is in place." on 2017-jun-06 the arjohuntleigh representative replaced middle pipe and lift arm on which the locking fitting for bath chair is positioned.However, this service was performed over 7 months before the issue was noticed by the customer.The involved device was under the arjohuntleigh service contract and its preventive maintenance was performed in 2017-dec-04.According to the information received, at that time no issue related to locking fitting was detected.Based on the collected information the involved device failed to meet the manufacturer's specification as the locking fitting was not mounted properly.At the time of event the device was used with a patient (customer of the facility).Please note that no adverse event occurred in this case.This complaint was reported to competent authorities based on initial information, however upon receiving additional information and clarification of the event circumstances, it was concluded that this incident does not meet the definition of mdr adverse event and will not be considered as such in the future.
|
|
Search Alerts/Recalls
|
|
|