This is filed to report the clip movement.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The first mitraclip (71103u159) was implanted medial, at a3/p3.The second clip delivery system (cds 71102u168) was advanced to the mitral valve.When advancing the clip to the left ventricle, the clip became caught on the chordae.Troubleshooting was performed and the clip was freed.There was no chordal damage.While retracting to the left atrium to reposition the clip, there was contact with the first clip implanted, and the first clip appeared to be loosened on one leaflet.The second clip was deployed to stabilize the first clip and for the mr reduction.Two clips were implanted, reducing mr to 1.It was noted that there was some shadowing from the guide during the procedure; however, this did not affect grasping and leaflet insertion.No additional information was provided.
|
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on the information provided the reported partial clip movement on the leaflet and device damaged by another device appear to be due to a combination of the medial prolapse, thickened leaflets and user technique/procedural circumstances.The poor image resolution issue was associated with some shadowing from the guide that was noted while visualizing the clip during the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|