Model Number VICM5_12.6
Device Problem Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/17/2017
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.No similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon inserted a 12.6mm vicm5_12.6 implantable collamer lens, -13.00 diopter, and the lens went into the eye upside down.The lens tore as it was removed during the initial surgery, with no apparent injury.The surgeon was unable to implant the backup lens due to the pupil constricted.The surgery was aborted and another same model/diopter lens was implanted on (b)(6) 2017.The reporter indicated the cause of the event was unknown.
 
Manufacturer Narrative
Device evaluation: the lens was returned dry, in a micro-centrifuge vial with residue on product.Visual observation found the lens missing a piece of haptic and residue on lens surface.The cartridge was returned in a device tray.Visual inspection of the cartridge found no visible damage to the device.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key7294660
MDR Text Key100932568
Report Number2023826-2018-00276
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/30/2020
Device Model NumberVICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2018
Initial Date FDA Received02/24/2018
Supplement Dates Manufacturer Received03/22/2018
Supplement Dates FDA Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
-
-